DELETE
SAMPLE
Cause
Lamp is defect.
Remedy
Replace the lamp.
20° Lamp defect !
Continue:
OPERATE
Please Recalibrate!
Continue:
Calibration:
SET UP
OPERATE
Cause
Lamp aging or loss of stored calibration
values.
Remedy
Re-calibrate.
Cause
The storage capacity is exhausted
(600 memory locations).
Remedy
If possible, transfer stored data to a PC.
Memory full!
Delete:
Escape:
DELETE
SAMPLE
Cause
Lamp is defect.
Remedy
Replace the lamp.
20° Lamp defect !
Continue:
OPERATE
Please Recalibrate!
Continue:
Calibration:
SET UP
OPERATE
Cause
Lamp aging or loss of stored calibration
values.
Remedy
Re-calibrate.
Cause
The storage capacity is exhausted
(600 memory locations).
Remedy
If possible, transfer stored data to a PC.