Symptom
After completion of the 10 sec.
preparation time, the
displaycontinues to show
"overflow"
.
Possible Cause and Remedy
Check connection of
measuring cable. If
unsuccessful, the
resistance of the sample
may be too high (>19,99
M).
Symptom
Fluctuating, unstable values are
displayed or considerable errors of
measurement occur.
Possible Cause and Remedy
Electric field effects can cause
this problem. A shielded low-
capacitance probe cable is
required.
Repair Work
Please contact the BYK-Gardner
Service Department.